Service manual
○
C
Copyright 2009 Agilent Technologies
Change 2
Change the Measurement Uncertainty value of OSC Level Accuracy Test (Page 69 to 70) to
the following information and appended TAR to this report.
OSC Level Accuracy Test
OSC Level Accuracy (Accessory: Nothing)
OSC
Level
Frequency Test Limits
Test Result Measurement
Uncertainty
TAR
1.0 V
1 kHz ± 101 mV
mV
± 0.4 mV 271
100 kHz ± 101 mV
mV
± 1.21 mV 84
0.5 V
40 Hz ± 51.0 mV
mV
± 0.06 mV 326
1 kHz ± 51.0 mV
mV
± 0.09 mV 204
10 kHz ± 51.0 mV
mV
± 0.09 mV 289
100 kHz ± 51.0 mV
mV
± 0.42 mV 65
1 MHz ± 51.3 mV
mV
± 5.2 mV 15
10 MHz ± 53.5 mV
mV
± 5.3 mV 20.2
15 MHz ± 54.8 mV
mV
± 5.3 mV 16.1
16 MHz ± 55.0 mV
mV
± 5.3 mV 15.3
50 MHz ± 63.5 mV
mV
± 5.4 mV 16.6
110 MHz ± 78.5 mV
mV
± 5.7 mV 18.2
0.25 V
1 kHz ± 26.0 mV
mV
± 0.06 mV 195.2
100 kHz ± 26.0 mV
mV
± 0.22 mV 80.4
0.125 V
1 kHz ± 13.5 mV
mV
± 0.04 mV 167.8
100 kHz ± 13.5 mV
mV
± 0.12 mV 79.2
0.064 V
1 kHz ± 7.4 mV
mV
± 0.01 mV 269.1
100 kHz ± 7.4 mV
mV
± 0.05 mV 78.9
0.016 V
1 kHz ± 2.6 mV
mV
± 0.004 mV 52.2
100 kHz ± 2.6 mV
mV
± 0.02 mV 46.0