Service manual

5 131
Power Supply Troubleshooting
TROUBLSHOOT THE FAN AND THE A50 DC-DC CONVERTER
5. Power Supply
Troubleshooting
TROUBLSHOOT THE FAN AND THE A50 DC-DC
CONVERTER
Perform the following procedure to troubleshoot the fan and the A50 DC-DC Converter.
1. Troubleshoot the Fan
1. Turn the analyzer power off.
2. Disassemble the rear panel.
3. Remove the fan power cable from the Motherboard A20J22.
4. Connect a DC power supply, a 10 kΩ resistance, and a oscilloscope to the fan power
cable using appropriate wire as shown in
Figure 5-9.
Figure 5-9 Fan Troubleshooting Setup
5. Turn the DC power supply on. Adjust the output voltage to +24V.
6. Check the fan is rotating. Check the FAN LOCK signal is as shown in Figure 5-9
If the fan is not rotating or the FAN LOCK signal is unexpected, replace the fan.
If these are good, the fan is verified.
7. Reconnect the fan power cable to the Motherboard A20J22.