Service manual
3 103
Adjustment
Frequency Reference Adjustment (Opt.1D5 only)
1. Chapter Title
2. Chapter Title 3. Adjustment
4. Chapter Title
5. Chapter Title
Frequency Reference Adjustment (Opt.1D5 only)
The purpose of this procedure is to adjust the signal frequency accuracy.
NOTE The adjustment program doesn’t include this adjustment. Perform this by manual
operation.
Required Equipment
Description Recommended Model
Frequency Counter 53181A or 5334B
Frequency Standard 5071A or 5061B
BNC(m)-BNC(m) Cable, 61 cm p/n 04294-61001
Dual Banana-BNC(f) Adapter p/n 1251-2277
Procedure
1. Connect the equipment as shown in Figure 3-16
Figure 3-16 Frequency Reference Adjustment Setup