Service manual

100 3
Adjustment
Impedance Measurement Correction Constants
Impedance Measurement Correction Constants
The purpose of this procedure is to adjust the relative impedance between the range
resistors.
Required Equipment
Description Recommended Model
Multimeter 3458A
100 Ω resistor p/n 04294-61001
Banana(m)-Banana(m) Cable 11058A, 2ea
BNC(f)-BNC(f) Adapter p/n 1250-1830, 4ea
BNC(m)-Banana(f) Adapter p/n 1250-1263, 4ea
Procedure
1. Run the adjustment program.
2. Choose the Adj_4TP_LOAD.
3. Connect the 100 Ω resistor to the multimeter as shown in Figure 3-14