Adjustment
OSC Level Correction Constants
1. Chapter Title
2. Chapter Title 3. Adjustment
4. Chapter Title
5. Chapter Title
OSC Level Correction Constants
The purpose of this procedure is to adjust the output oscillation level.
Required Equipment
Description Recommended Model
4TP OPEN termination 42090A
Procedure
1. Run the adjustment program.
2. Choose the Adj_OSC_LEVEL.
3. Connect the equipment as shown in Figure 3-13
Figure 3-13 OSC Level Correction Constants Setup
4. Follow the adjustment program instruction to update the correction constants.