Service manual

3 97
Adjustment
OSC Level Monitor Correction Constants
1. Chapter Title
2. Chapter Title 3. Adjustment
4. Chapter Title
5. Chapter Title
OSC Level Monitor Correction Constants
The purpose of this procedure is to adjust the OSC level monitor.
Required Equipment
Description Recommended Model
Multimeter 3458A
Interface Box p/n 04294-61002
BNC(m)-BNC(m) Cable, 61 cm p/n 8120-1839
Dual Banana-BNC(f) Adapter p/n 1251-2277
Power Meter 437B, 438A,
E4418A/B,
or E4419A/B
Power Sensor 8482A
Power Splitter 11667A
50 Ω termination, N(m) 909C Opt.012
Procedure
1. Run the adjustment program.
2. Choose the Adj_OSC_MON.
3. Connect the equipment as shown in Figure 3-11