Service manual

3 95
Adjustment
DC Bias Monitor Correction Constants
1. Chapter Title
2. Chapter Title 3. Adjustment
4. Chapter Title
5. Chapter Title
Figure 3-8
Figure 3-8 DC Bias Monitor Correction Constants Setup 2
6. Click OK, then wait till the measurement finish.
7. Disconnect the dual banana plug from the multimeter, then connect it as shown in
Figure 3-9
Figure 3-9 DC Bias Monitor Correction Constants Setup 3
8. Follow the adjustment program instruction to update the correction constants.