Service manual

C
Copyright 2009 Agilent Technologies
Agilent 4294A PRECISION IMPEDANCE ANALYZER Service Manual
Manual Change
Agilent Part No. N/A
May 2009
Change 1
Change the Measurement Uncertainty value of Frequency Accuracy Test (Page 68) to the
following information and appended TAR to this report.
Frequency Accuracy Test
Without Option 1D5
Frequency Test Limit Test Result
1
Measurement
Uncertainty
TAR
10MHz ±200 Hz Hz ± 0.3 Hz 890
With Option 1D5
Frequency Test Limit Test Result
1
Measurement
Uncertainty
TAR
10MHz ±1.3 Hz Hz ± 0.2 Hz 890