Instruction manual
234 Chapter 8 
Analysis and Processing of Result
Define the Peak
“Search the maximum/minimum peak” on page 228. It should be noted, however, that a 
definition must be set in advance by using the 
PEAK DEF DX key and PEAK DEF DY key 
to ensure that the target peak is searched. To ensure that all peaks are detected, perform 
peak search after setting the peak 
DY value to zero using the PEAK DEF DY.
Step 3.  Press the 
[Search] key to display the Search menu. 
Step 4. Press the 
PEAK key to display the Peak menu.
Step 5. Press the PEAK DEF MENU key to display the Peak Definition menu.
Step 6. Press the 
MKR ® PEAK DELTA key to enter the peak sharpness at the main marker 
position into the peak definition of 
DX and DY .
NOTE When using the MKR ® PEK DELTA key to define the peak, the sharpness of the left side 
(where the sweep parameter value is smaller) of the peak at the main marker position is 
used to define DX and DY.
With both the D marker and the main marker displayed on the screen, if you press the 
MKR 
® PEK DELTA
, the sweep parameter value (absolute value) of the difference between the 
main marker and the D marker is assigned to the definition of DX; correspondingly, the 
measurement parameter value (absolute value) of this difference is assigned to the 
definition of DY.