Instruction manual

Chapter 6 135
Calibration
B. Calibration for Four-Terminal Pair, 1-m Extension
6. Calibration
B. Calibration for Four-Terminal Pair, 1-m Extension
This section describes the calibration procedures to follow when the four-terminal pair port
of the Agilent 4294A is extended with a 1-meter-long four-terminal pair configuration and
the device is connected to the cable’s end with a minimal connection.
Figure 6-2 Calibration for Four-Terminal Pair, 1-m Extension
This calibration procedure is applied to the following connections:
The 16048G (1-m Cable) is connected to the Agilent 4294A and the device is
connected to the cable’s end with a test fixture.
The 16048G (1-m Cable) is connected to the Agilent 4294A and the 16452A (Liquid
Test Fixture) is connected to the end of the cable.
The 16334A (Tweezer Type Test Fixture) is connected to the Agilent 4294A.
The 16451B (Dielectric Test Fixture) is connected to the Agilent 4294A.
Step 1. Make sure that the adapter selection is set to “Four-terminal pair, 1-meter (
4TP 1M).”
See “Adapter Setting” on page 67 for adapter selection.
Step 2. Perform the fixture compensation in accordance with Fixture Compensation” on page