Specifications

Chapter 6 97
Retrieving Measurement Results
Retrieving the Measurement Results for Measurement Parameters 1
through 4
6. Retrieving Measurement
Results
Example 6-5 Using the :READ? Command to Retrieve Measurement Results (read.bas)
10 DIM Para1$[9],Para2$[9]
20 REAL Res1,Res2,Imon
30 INTEGER Point,Stat
40 ASSIGN @Agt4287a TO 717
50 Point=1
60 OUTPUT @Agt4287a;":FORM ASC"
70 !
80 OUTPUT @Agt4287a;":DISP:TEXT1:CALC1 ON"
90 OUTPUT @Agt4287a;":DISP:TEXT1:CALC2 ON"
100 OUTPUT @Agt4287a;":DISP:TEXT1:CALC3 OFF"
110 OUTPUT @Agt4287a;":DISP:TEXT1:CALC4 OFF"
120 OUTPUT @Agt4287a;":DISP:TEXT1:CALC11 ON"
130 OUTPUT @Agt4287a;":DISP:TEXT1:CALC12 OFF"
140 !
150 OUTPUT @Agt4287a;":CALC:PAR1:FORM?"
160 ENTER @Agt4287a;Para1$
170 OUTPUT @Agt4287a;":CALC:PAR2:FORM?"
180 ENTER @Agt4287a;Para2$
190 !
200 OUTPUT @Agt4287a;":SOUR:LIST:STAT OFF"
210 OUTPUT @Agt4287a;":SOUR:LIST:POIN ";Point
220 OUTPUT @Agt4287a;":CALC:COMP OFF"
230 OUTPUT @Agt4287a;":SOUR:LIST:RDC OFF"
240 !
250 ! Trigger source setting
260 !
270 OUTPUT @Agt4287a;":ABOR"
280 OUTPUT @Agt4287a;":TRIG:SOUR EXT"
290 !
300 ! Triggering and data read
310 !
320 OUTPUT @Agt4287a;":READ?"
330 PRINT "Waiting for External Trigger!"
340 ENTER @Agt4287a;Stat,Res1,Res2,Imon
350 !
360 ! Display results
370 !
380 PRINT "### Result ###"
390 PRINT "Meas. Status:";Stat,Para1$&":";Res1,Para2$&":";Res2,"Imon
:";Imon
400 !
410 END