Specifications

84 Chapter 6
Retrieving Measurement Results
Internal Data Processing
R
dc
compensation coefficients and compensation data
When the instrument performs R
dc
measurement (DC measurement), it uses a specific set
of compensation coefficients (real numbers) that are different from those used for
impedance measurement. This means that the standard measurement data (real numbers)
necessary for calculating the compensation coefficients is measured and retained
separately from those for impedance measurement.
The instrument has three compensation coefficients, A, B, and C, for R
dc
measurement. To
set or retrieve one of these coefficients, use one of the following commands:
:DATA:RCMP{1-3} on page 266
To obtain the OPEN or SHORT measurement data for calculating the compensation
coefficients for R
dc
measurement, use one of the following commands:
:DATA:RCMD{1-2}? on page 266
Data number
*1
Data to retrieve
1R
dc
Calibration Coefficient A
2R
dc
Calibration Coefficient B
3R
dc
Calibration Coefficient C
*1.Use this number as the suffix to the command.
Data number
*1
Data to retrieve
1
OPEN measurement data for calculating the compensation coefficients for
R
dc
measurement
2
SHORT measurement data for calculating the compensation coefficients for
R
dc
measurement
*1.Use this number as the suffix to the command.