Specifications

Chapter 6 83
Retrieving Measurement Results
Internal Data Processing
6. Retrieving Measurement
Results
R
dc
Display Data
The R
dc
Display Data is the data obtained by performing R
dc
-specific calibration and
compensation on the results of R
dc
measurement. To retrieve this data, use the following
command:
:DATA:RDC? on page 267
Calibration coefficients and calibration data for R
dc
measurement
When the instrument performs R
dc
measurement (DC measurement), it uses a specific set
of calibration coefficients (real numbers) that are different from those used for impedance
measurement. This means that the standard measurement data (real numbers) necessary for
calculating the calibration coefficients are measured and retained separately from those for
impedance measurement.
The instrument has three calibration coefficients, A, B, and C, for R
dc
measurement. To set
or retrieve one of these coefficients, use one of the following commands:
:DATA:RCCO{1-3} on page 265
To obtain the OPEN, SHORT, or LOAD measurement data for calculating the calibration
coefficients for R
dc
measurement, use one of the following commands:
:DATA:RCAD{1-3}? on page 265
Data number
*1
Data to retrieve
1R
dc
Calibration Coefficients
2R
dc
Calibration Coefficients
3R
dc
Calibration Coefficients
*1.Use this number as the suffix to the command.
Data number
*1
Data to retrieve
1
OPEN measurement data for calculating the calibration coefficients for R
dc
measurement
2
SHORT measurement data for calculating the calibration coefficients for R
dc
measurement
3
LOAD measurement data for calculating the calibration coefficients for R
dc
measurement
*1.Use this number as the suffix to the command.