Specifications
82 Chapter 6
Retrieving Measurement Results
Internal Data Processing
Current/Voltage Monitor Arrays
The Current/Voltage Monitor Arrays contain the real numbers to be displayed as the
monitored values of the test signal voltage and current levels. Each value contained in
these arrays is calculated based on the test signal level setting and the impedance value
determined by performing port extension and compensation on the Data Array.
To retrieve the Voltage or Current Monitor Array, use one of the following commands:
Command Description
:DATA:VMON? on page 267 Retrieves the Voltage Monitor Array.
:DATA:IMON? on page 263 Retrieves the Current Monitor Array.