Specifications

78 Chapter 6
Retrieving Measurement Results
Internal Data Processing
Internal data arrays
The instrument has a number of internal data arrays/sets that can be read and/or written and
contain either real or complex numbers (Table 6-1).
NOTE Each table has its own calibration/compensation-related internal data arrays and data sets
(including Calibration/Compensation Coefficient Arrays, Calibration/Compensation Data
Arrays, R
dc
Calibration/Compensation Coefficients, and R
dc
Calibration/Compensation
Data). When you perform command-based read/write operations on these data arrays and
sets, the instrument assumes that the active table is the target.
NOTE A complex-number array has two values (real and imaginary parts) for each measurement
point, thus containing Nopยด2 elements in all, where Nop represents the number of
measurement points.
Table 6-1 Internal data arrays/sets
Name of data array/set Read Write Data format
Data array Allowed Not allowed Complex number
Display data arrays Allowed Not allowed Real number
Calibration Data Arrays Allowed Not allowed Complex number
Calibration Coefficient Arrays Allowed Allowed Complex number
Compensation Data Arrays Allowed Not allowed Complex number
Compensation Coefficient
Arrays
Allowed Allowed Complex number
Current/Voltage Monitor Array Allowed Not allowed Real number
R
dc
Display Data Allowed Not allowed Real number
R
dc
Calibration Data Allowed Not allowed Real number
R
dc
Calibration Coefficients Allowed Allowed Real number
R
dc
Compensation Data Allowed Not allowed Real number
R
dc
Compensation Coefficients Allowed Allowed Real number