Specifications
66 Chapter 5
Starting Measurement Cycle (Triggering) and Detecting End of Measurement
Starting a New Measurement Cycle (Triggering)
Starting a New Measurement Cycle (Triggering)
Trigger system
The trigger system is responsible for such tasks as detecting the start of a measurement
cycle (triggering) and controlling the pre-measurement delay time. As shown in Figure 5-1,
the trigger system has three states: Idle, Waiting for Trigger, and Measurement.
Figure 5-1 Trigger system
The following subsections describe each state and how the trigger system switches among
the three states.
Idle state
The trigger system is put into Idle state (arrow e in Figure 5-1) when you power on the
instrument or issue one of the following commands:
• *RST on page 206
• :ABOR on page 209
The trigger system is put into Waiting for Trigger state (arrow a in Figure 5-1) when you
activate it with one of the following commands:
• :INIT on page 287
• :INIT:CONT on page 287 (“ON” specified)