Specifications
52 Chapter 4
Preparing for Accurate Measurement
Running the Compensation Process
Running the Compensation Process
NOTE You must turn on the calibration function before using it. In other words, when the
calibration function is off, you cannot perform any of the compensation related operations,
including measuring the data for calculating the compensation coefficients, calculating the
compensation coefficients, turning on or off the compensation function, and retrieving or
setting the compensation coefficients. However, you can configure the compensation kit
even when the calibration function is off.
Setting compensation kit values
You can either use the 4287A's predefined compensation kit values or directly enter other
compensation kit values. To choose how the compensation kit values are defined, use the
following command:
• :CORR2:CKIT on page 247
If you opt to enter the values, you need to specify how the compensation kit itself is
defined and then enter the definitions.
Choosing definition method (Fixed or Point-by-Point)
• :CORR2:CKIT:LIST on page 247
Entering definitions
To define the OPEN and SHORT standards, use the commands listed in the following
table:
Measuring data for calculating compensation coefficients
To measure the data for calculating the compensation coefficients, use the following
command:
• :CORR2:COLL on page 252
The command shown above returns the data for calculating the compensation coefficients
for both impedance and R
dc
measurements. If you want to obtain the data for either
impedance or R
dc
measurement alone, use one of the following commands instead:
• :CORR2:COLL:RF on page 254
• :CORR2:COLL:DC on page 253
NOTE Issuing the :CORR2:COLL, :CORR2:COLL:RF, or :CORR2:COLL:DC command
does not trigger the instrument to begin measuring the data. This means that if the trigger
Impedance value R
dc
value
OPEN :CORR2:CKIT:STAN1:LIST on page 249 :CORR2:CKIT:STAN1:DC on page 248
SHORT :CORR2:CKIT:STAN2:LIST on page 251 :CORR2:CKIT:STAN2:DC on page 250