Specifications

362 Appendix G
Initial Settings
Initial Settings, Settings that can be Saved/Recalled, Settings that can be
Backed Up
Setting items
Initial settings
(factory settings)
Reset
Save/
Recall
Backup
Available
method
to set
:SYST:PRES *RST
Target table in measurement of data for calibration coefficient
calculation (all tables/active table only)
*1
Active table only No effect ¬´m Front
Automatically saves the state into auto-recall file (autorec.sta)
when the calibration coefficients are calculated
*2
No check No effect ¬´m Both
Calibration kit selection (7 mm standard/user defined) 7 mm standard ¬¬m ´ Both
User
defined
calibration
kit
Impedance definition method
(fixed/point-by-point)
Fixed ¬¬m ´ Both
Open Equivalent parallel conductance (G) 0 [S] ¬¬m ´ Both
Equivalent parallel capacitance (Cp) 0 [F] ¬¬m ´ Both
Dc conductance 0 [S] ¬¬m ´ Both
Offset delay time 0 [s] ¬¬m ´ Both
Short Equivalent series resistance (Rs) 0 [W] ¬¬m ´ Both
Equivalent series inductance (Ls) 0 [H] ¬¬m ´ Both
Dc resistance 0 [W] ¬¬m ´ Both
Offset delay time 0 [s] ¬¬m ´ Both
Load Definition impedance parameter type Rs and Ls ¬¬m ´ Both
Equivalent series resistance (Rs) 50 [W] ¬¬m ´ Both
Equivalent series inductance (Ls) 0 [H] ¬¬m ´ Both
Dc resistance 50 [W] ¬¬m ´ Both
Offset delay time 0 [s] ¬¬m ´ Both
*1. When data measurement is performed by GPIB command, the target table is always ‘active table only’ regardless
of this setting.
*2. When calculation of the calibration coefficients is performed by GPIB command, the state is not saved into
auto-recall file (autorec.sta) regardless of this setting.