Specifications
356 Appendix F
List of Responses to Measurement Failure
Responses to Measurement Failure
Responses to Measurement Failure
When any of the events listed below is detected, the instrument responds by displaying the
corresponding notification on the LCD screen, outputting the corresponding data through
GPIB, and activating the corresponding handler interface signal, as shown in Table F-1.
• Overload
(This failure condition is detected when the instrument fails in its ranging
performance; for example, when the DUT is detached during measurement.)
•R
dc
out of range
(This failure condition is detected when the R
dc
measurement result falls outside of the
limit range.)
• Overload and R
dc
out of range
Table F-1 List of instrument responses to overload and R
dc
out of range conditions
Notification on LCD screen GPIB output
Activated handler
interface signal(s)
Measurement results
Comparator
sorting
results
Measurement
status
Measured
values
*1
Comparator
sorting
results
Parameters
1 through 4
Test signal
level
monitoring
R
dc
Overload
*2
9.9E37 9.9E37
Measured
values
*3
14
19.9E37
14
/OVLD
R
dc
out of
range
Measured
values
*4
Measured
values
*4
Measured
values
*5
(highlighted
in red)
2
Measured
values
*6
/RDC_OUT_OF_RANGE
Overload
*2
and
R
dc
out of
range
9.9E37 9.9E37
Measured
values
*5
(highlighted
in red)
39.9E37
/OVLD
/RDC_OUT_OF_RANGE
*1. All of the measured values including the measurement results for parameters 1 through 4, test signal level moni-
toring, and R
dc
measurement.
*2. In the instrument status area (near the bottom of the screen), an “Ovld” indicator appears in red.
*3. The instrument never detects an overload condition during R
dc
measurement.
*4. Measured values are displayed as usual, but they can be incorrect due to imperfect contact.
*5. Measured values are displayed as usual, but they are highlighted in red.
*6. Measured values are retrieved as usual, but they can be incorrect due to imperfect contact.