Specifications
Chapter 3 35
Specifying Measurement Conditions
Sample Program
3. Specifiying Measurement
Conditions
620 !
630 FOR I=1 TO 4
640 IF Para$(I)="OFF" THEN
650 OUTPUT @Agt4287a;":DISP:TEXT1:CALC"&VAL$(I)&" OFF"
660 ELSE
670 OUTPUT @Agt4287a;":DISP:TEXT1:CALC"&VAL$(I)&" ON"
680 OUTPUT @Agt4287a;":CALC:PAR"&VAL$(I)&":FORM "&Para$(I)
690 IF Dev$(I)="OFF" THEN
700 OUTPUT @Agt4287a;":CALC:PAR"&VAL$(I)&":EXPR:STAT OFF"
710 ELSE
720 OUTPUT @Agt4287a;":CALC:PAR"&VAL$(I)&":EXPR:STAT ON"
730 OUTPUT @Agt4287a;":CALC:PAR"&VAL$(I)&":EXPR:NAME "&Dev$(I)
740 OUTPUT @Agt4287a;":CALC:PAR"&VAL$(I)&":EXPR:CENT ";Cent(I)
750 END IF
760 END IF
770 NEXT I
780 !
790 ! Measurement Point(s)
800 !
810 OUTPUT @Agt4287a;":SOUR:LIST:POIN ";Point_no
820 OUTPUT @Agt4287a;":SOUR:LIST:STAT "&List$
830 !
840 ! Test Signal Level Monitor Result Display
850 !
860 OUTPUT @Agt4287a;":DISP:TEXT1:CALC11 "&Imon_disp$
870 OUTPUT @Agt4287a;":DISP:TEXT1:CALC12 "&Vmon_disp$
880 !
890 ! Rdc Measurement
900 !
910 OUTPUT @Agt4287a;":SOUR:LIST:RDC "&Rdc$
920 OUTPUT @Agt4287a;":CALC:COMP:RDC:LIM ";Rdc_l_lim;",";Rdc_u_lim
930 !
940 ! Displayed Parameter in List Measurement Display
950 !
960 FOR I=1 TO 4
970 OUTPUT @Agt4287a;":DISP:TEXT2:LAB"&VAL$(I)&" "&List_lab$(I)
980 NEXT I
990 !
1000 ! Title Display Area
1010 !
1020 OUTPUT @Agt4287a;":DISP:TEXT10 "&Title_disp$
1030 OUTPUT @Agt4287a;":DISP:TEXT10:DATA """&Title$&""""
1040 !
1050 ! Clock (Data/Time) Display
1060 !
1070 IF Clock_disp$="OFF" THEN
1080 OUTPUT @Agt4287a;":DISP:TEXT11 OFF"
1090 ELSE
1100 OUTPUT @Agt4287a;":DISP:TEXT11 ON"
1110 OUTPUT @Agt4287a;":DISP:TEXT11:MODE "&Clock_disp$
1120 END IF
1130 END