Specifications

Appendix E 347
4286A vs. 4287A GPIB Commands Correspondence Table
4286A vs. 4287A GPIB Commands Correspondence Table
2. Chapter Title 3. Chapter Title
4. Chapter Title
5. Chapter Title
E. 4286A vs. 4287A GPIB
Commands Correspondence
Tab l e
Compensation Select Compensation Kit None. :CORR2:CKIT on page 247
Set User Definition Compensation
Kit Label
:SENS:CORR2:CKIT:LAB None.
Save User Definition
Compensation Kit
:SENS:CORR2:CKIT:SAVE None.
(Saved with the hardware settings.)
Define Each
Standard of
User Definition
Compensation
Kit
Select How to
Define Impedance
Value (for each
fixed point and
measurement
point)
None. :CORR2:CKIT:LIST on page 247
Define Open :SENS:CORR2:CKIT:STAN1:G
:SENS:CORR2:CKIT:STAN1:C
:CORR2:CKIT:STAN1:LIST on
page 249
:CORR2:CKIT:STAN1:DC on
page 248
Define Short :SENS:CORR2:CKIT:STAN2:R
:SENS:CORR2:CKIT:STAN2:L
:CORR2:CKIT:STAN2:LIST on
page 251
:CORR2:CKIT:STAN2:DC on
page 250
Define Load :SENS:CORR2:CKIT:STAN3:R
:SENS:CORR2:CKIT:STAN3:L
None.
(Cannot perform load compensation.)
Select Definition Value of User
Definition Compensation Kit
:SENS:CORR2:CKIT:STAN1
:SENS:CORR2:CKIT:STAN2
:SENS:CORR2:CKIT:STAN3
None.
Measure Compensation Coefficient
Calculation Data
:SENS:CORR2:COLL :CORR2:COLL on page 252
:CORR2:COLL:RF on page 254
:CORR2:COLL:DC on page 253
Select Measurement Points for
Calibration Coefficient Calculation
Data
:SENS:CORR2:COLL:FPO None.
(There is no mode for measuring data
with fixed measurement points.)
Calculate Compensation
Coefficient
:SENS:CORR2:COLL:SAVE :CORR2:COLL:SAVE on page 255
Set Open Compensation Function
ON/OFF
:SENS:CORR2:OPEN :CORR2:COLL:OPEN on page 254
Set Short Compensation Function
ON/OFF
:SENS:CORR2:SHOR :CORR2:COLL:SHOR on page 255
Set Load Compensation Function
ON/OFF
:SENS:CORR2:LOAD None.
(Cannot perform load compensation.)
Fixture
Selection
Fixture Selection :SYST:FIXT :CORR2:FIXT on page 256
Set User-created
Fixture
Amount
Compensated
(Electrical
Length)
:SYST:FIXT:DIST :CORR2:FIXT:EDEL:DIST on
page 257
Name :SYST:FIXT:LAB :CORR2:FIXT:LAB on page 258
Save :SYST:FIXT:SAVE None.
(Saved with the hardware settings.)
Function Item to set/run
GPIB command
4286A 4287A