Specifications

346 Appendix E
4286A vs. 4287A GPIB Commands Correspondence Table
4286A vs. 4287A GPIB Commands Correspondence Table
Calibration Select Calibration Kit :SENS:CORR1:CKIT :CORR1:CKIT on page 231
Set User Definition Calibration Kit :SENS:CORR1:CKIT:LAB None.
Save User Definition Calibration
Kit
:SENS:CORR1:CKIT:SAVE None.
(Saved with the hardware settings.)
Define Each
Standard of
User Definition
Calibration Kit
Select How to
Define Impedance
Value (for each
fixed point and
measurement
point)
None. :CORR1:CKIT:LIST on page 232
Define Open :SENS:CORR1:CKIT:STAN1:G
:SENS:CORR1:CKIT:STAN1:C
:CORR1:CKIT:STAN1:LIST on
page 235
:CORR1:CKIT:STAN1:DC on
page 233
:CORR1:CKIT:STAN2:LIST on
page 238
Define Short :SENS:CORR1:CKIT:STAN2:R
:SENS:CORR1:CKIT:STAN2:L
:CORR1:CKIT:STAN2:LIST on
page 238
:CORR1:CKIT:STAN2:DC on
page 236
:CORR1:CKIT:STAN2:EDEL on
page 237
Define Load :SENS:CORR1:LOAD:TYPE
:SENS:CORR1:CKIT:STAN3:R
:SENS:CORR1:CKIT:STAN3:L
:SENS:CORR1:CKIT:STAN3:Q
:SENS:CORR1:CKIT:STAN3:F
:CORR1:CKIT:STAN3:FORM on
page 241
:CORR1:CKIT:STAN3:LIST on
page 242
:CORR1:CKIT:STAN3:DC on
page 239
:CORR1:CKIT:STAN3:EDEL on
page 240
Measure Calibration Coefficient
Calculation Data
:SENS:CORR1:COLL :CORR1:COLL on page 244
:CORR1:COLL:RF on page 246
:CORR1:COLL:DC on page 245
Select Measurement Points for
Calibration Coefficient Calculation
Data
:SENS:CORR1:COLL:FPO None.
(There is no mode for measuring data
with fixed measurement points.)
Check Calibration Function
ON/OFF
:SENS:CORR1? :CORR1? on page 231
Calibrate Calibration Coefficient
and Turn On Calibration Function
:SENS:CORR1:COLL:SAVE :CORR1:COLL:SAVE on page 246
Set Extension of Port :SENS:CORR1:EDEL:STAT
:SENS:CORR1:EDEL
None.
(The same function can be implemented
with the electrical length setting of
fixture selection.)
Function Item to set/run
GPIB command
4286A 4287A