Specifications
32 Chapter 3
Specifying Measurement Conditions
Sample Program
Sample Program
Example 3-1 shows a sample program that demonstrates how to configure measurement
conditions. You can find the source file of this program, named setup.bas, on the sample
program disk.
This program resets the instrument and then configures it with the following conditions:
• Active table number: 3
• Point number for single-point measurement: 1
• Measurement point setup table:
• Settings of Measurement Parameters 1 through 4:
• List measurement: On
• Display parameters for list measurement screen
Display item 1: Measuring frequency
Display item 2: Measurement parameter (Ls)
Display item 3: Measurement parameter (Rs)
Display item 4: Measurement parameter (Q)
• Result of test signal current level monitoring: On
• Result of test signal voltage level monitoring: On
•R
dc
measurement: On
•R
dc
measurement limit range
Lower limit: -5 W
Upper limit: 5 W
• Title display area: On (title: "Example 3-1")
• Time and date display: On (auto update)
Point number Measuring frequency Averaging factor Signal source level
1 1 MHz 2 -10 dBm
2 10 MHz 2 -10 dBm
3 100 MHz 2 -10 dBm
4 1 GHz 2 -10 dBm
Parameter Display
Deviation mode
On/off How parameter is
displayed
Reference
value
Measurement parameter 1 Ls On On Percentage 100 nH
Measurement parameter 2 Rs On On Absolute value 100 nH
Measurement parameter 3 Q On Off —— ——
Measurement parameter 4 —— Off —— —— ——