Specifications
Chapter 16 311
Command Reference
Service command
16. Command Reference
Service command
This section describes the command for service engineers. If you use the command shown
in this section, accuracy of measurement is not guaranteed.
:SOUR:LIST:RDC:AVER
Syntax :SOURce:LIST:RDC:AVERage <numeric>
:SOURce:LIST:RDC:AVERage?
Description Sets the averaging factor of the R
dc
measurement.
For maximum reduction of measurement variations by ac line noise at measurements for
service, set the averaging factor to following number.
When ac line frequency is 50 Hz: 2000
When ac line frequency is 60 Hz: 1667
Parameters
If your specified parameter is beyond the valid range, the instrument ignores the parameter
and uses either the maximum or minimum value depending on whether the parameter is
greater than the upper limit or smaller than the lower limit.
Query response {numeric}<newline><^END>
Equivalent key
sequence
No equivalent keys are available on the front panel.
<numeric>
Description Averaging factor
Range 1 to 6000
Default 128
Resolution 1