Specifications
Chapter 16 297
Command Reference
:SOUR:LIST:STAT
16. Command Reference
:SOUR:LIST:STAT
Syntax :SOURce:LIST:STATe {ON|OFF|1|0}
:SOURce:LIST:STATe?
Description Sets the measurement points to be used during measurement.
List measurement The instrument measures the DUT’s characteristics at all points
defined in the active table.
Single-point
measurement The instrument measures the DUT’s characteristics only at the point
specified by issuing the :SOUR:LIST:POIN command.
Parameters
Query response {1|0}<newline><^END>
See also :SOUR:LIST:POIN on page 295
Equivalent key
sequence
[Meas View]
:SOUR:LIST:TABL
Syntax :SOURce:LIST:TABLe <numeric>
:SOURce:LIST:TABLe?
Description Specifies which table to use as the active table.
Parameters
If your specified parameter is beyond the valid range, the instrument ignores the parameter
and uses either the maximum or minimum value depending on whether the parameter is
greater than the upper limit or smaller than the lower limit.
Query response {numeric}<newline><^END>
Equivalent key
sequence
[Stml Select] - TABLE No.
Description
ON or 1 Instructs instrument to perform list measurement
OFF or 0 (initial value) Instruct instrument to perform single-point measurement
<numeric>
Description Number of the table that should be the active table
Range 1 to 8
Default 1