Specifications
296 Chapter 16
Command Reference
:SOUR:LIST:RDC
:SOUR:LIST:RDC
Syntax :SOURce:LIST:RDC {ON|OFF|1|0}
:SOURce:LIST:RDC?
Description Specifies whether to perform R
dc
measurement. This setting is retained even if the active
table is changed.
Parameters
Query response {1|0}<newline><^END>
Equivalent key
sequence
[Prmtr] - RDC MEAS
:SOUR:LIST:RDC:OFSC
Syntax :SOURce:LIST:RDC:OFSCancel {ON|OFF|1|0}
:SOURce:LIST:RDC:OFSCancel?
Description Turns ON/OFF the offset cancel function for the R
dc
measurement.
NOTE Turn ON this setup before measuring calibration/compensation data.
Parameters
Query response {1|0}<newline><^END>
Corresponding key
[System] - SERVICE MENU - RDC OFS CANCEL
:SOUR:LIST:SIZE?
Syntax :SOURce:LIST:SIZE?
Description Returns the number of measurement points defined in the active table. (Query only)
Query response {numeric}<newline><^END>
Equivalent key
sequence
No equivalent keys are available on the front panel.
Description
ON or 1 Instructs instrument to perform R
dc
measurement (turns on R
dc
measurement function)
OFF or 0 (initial value) Instructs instrument to NOT perform R
dc
measurement (turns
off R
dc
measurement function)
Description
ON or 1 Turns ON the offset cancel function.
OFF or 0 (initial value) Turns OFF the offset cancel function.