Specifications

Chapter 16 295
Command Reference
:SOUR:LIST:CLE
16. Command Reference
:SOUR:LIST:CLE
Syntax :SOURce:LIST:CLEar
Description Resets all of the measurement point setup tables to the factory default settings, regardless
of the active table settings. (No query)
Equivalent key
sequence
[Alt] - [2] - [8] - [2] (when the measurement point setup display is selected by
[Setup View])
:SOUR:LIST:POIN
Syntax :SOURce:LIST:POINt <numeric>
:SOURce:LIST:POINt?
Description Specifies the measurement point for single-point measurement. This setting is retained
even if the active table is changed. If the specified point number exceeds the maximum
point number allowed in the new active table, however, the instrument will perform
single-point measurement using that maximum number instead of the previously specified
number.
This setting is reset when settings in the measurement point setup table are changed by
execution of the :SOUR:LIST or :SOUR:LIST:CLE command.
Parameters
If your specified parameter is beyond the valid range, the instrument ignores the parameter
and uses either the maximum or minimum value depending on whether the parameter is
greater than the upper limit or smaller than the lower limit.
Query response {numeric}<newline><^END>
See also :SOUR:LIST on page 294
:SOUR:LIST:CLE on page 295
:SOUR:LIST:STAT on page 297
Equivalent key
sequence
[Stml Select] - POINT No.
<numeric>
Description Point number for single-point measurement
Range 1 to number of points (up to 32)
Default 1
Resolution 1