Specifications
294 Chapter 16
Command Reference
:SOUR:LIST
:SOUR:LIST
Syntax :SOURce:LIST <nop>,<freq 1>,<ave 1>,<pow 1>,...,<freq N>,<ave N>,<pow N>
:SOURce:LIST?
where N represents the number of points contained in the active table (this number is
specified with the <nop> parameter).
Description Sets up the active table by specifying the measurement frequency values, averaging
factors, and oscillator levels for all measurement points.
Parameters
where N represents an integer between 1 and N.
If your specified parameter is beyond the valid range, the instrument ignores the parameter
and uses either the maximum or minimum value depending on whether the parameter is
greater than the upper limit or smaller than the lower limit.
Query response {nop},{freq 1},{ave 1},{pow 1},...,{freq N},{ave N},{pow N}<newline><^END>
See also :SOUR:LIST:SIZE? on page 296
:SOUR:UNIT on page 298
Equivalent key
sequence
[Setup View] (Measurement point setup display)
<nop> <freq n> <ave n>
Description Number of
measurement points
(N)
Measurement
frequency
at nth point
Averaging factor
at nth point
Range 1 to 32 1E6 to 3000E6 1 to 100
Default 1 1E6 1
Unit ———— Hz ————
Resolution 1 100E3 1
<pow n>
Description Oscillator level at nth point
Unit to apply
to setting
*1
dBm mA mV
Range -40 to 1 0.1 to 10 5 to 502
Default -13 2 100
Unit dBm mA mV
Resolution
0.1 dB
*2
*1. Use the :SOUR:UNIT command to select the unit for the setting.
*2.
The value is converted to dBm and then rounded off in resolutions of 0.1dB
when mA or mV is selected as the unit of oscillator level.