Specifications
Chapter 16 281
Command Reference
:FETC?
16. Command Reference
The bin sort result is not returned if the comparator is off.
NOTE If the measurement failed because of overload (that is, the measurement status is 1 or 3),
the command returns 9.9E37 for the measurement results of parameters 1 through 4, the
test signal monitor results, and the R
dc
measurement result; the command returns 14 for the
comparator result.
For list measurement
{stat 1},{par1 1},{par2 1},{par3 1},{par4 1},{Imon 1},{Vmon 1},{Rdc 1},...,
{stat N},{par1 N},{par2 N},{par3 N},{par4 N},{Imon N},{Vmon N},{Rdc N},
{comp}<newline><^END>
where N represents the number of points defined in the active table while n represents an
integer between 1 and N.
The command returns data at all measurement points in the active table. The returned data
consists of the following items:
{stat n}: Measurement status at measurement point number n
{par1 n}: Parameter 1 measurement result at measurement point number n
{par2 n}: Parameter 2 measurement result at measurement point number n
{par3 n}: Parameter 3 measurement result at measurement point number n
{par4 n}: Parameter 4 measurement result at measurement point number n
{Imon n}: Result of monitoring test signal current level at measurement point
number n
{Vmon n}: Result of monitoring test signal voltage level at measurement point
number n
{Rdc n}: Result of R
dc
measurement
{comp}: Bin sort result
NOTE The result of R
dc
measurement is returned N times. However, all of the results of R
dc
measurement have the same value because R
dc
is measured only once.
The bin sort result is finally returned only once.
See also :SOUR:LIST:STAT on page 297
:DISP:TEXT1:CALC{1-4} on page 270
:DISP:TEXT1:CALC{11-12} on page 272
:SOUR:LIST:RDC on page 296
:CALC:COMP on page 211
:READ? on page 293
*TRG on page 208
Equivalent key
sequence
No equivalent keys are available on the front panel.