Specifications
270 Chapter 16
Command Reference
:DISP:TEXT1:CALC{1-4}
:DISP:TEXT1:CALC{1-4}
Syntax :DISPlay[:WINDow]:TEXT1:CALCulate{1|2|3|4}[:STATe] {ON|OFF|1|0}
:DISPlay[:WINDow]:TEXT1:CALCulate{1|2|3|4}[:STATe]?
Description Turns on or off the screen display of the measurement results for each of the parameters 1
(:CALC1) through 4 (:CALC4).
Parameters
Query response {1|0}<newline><^END>
Equivalent key
sequence
[Display] - SETUP MEAS DISP - {PRMTR-1|PRMTR-2|PRMTR-3|PRMTR-4} - DISPLAY
:DISP:TEXT1:CALC{1-4}:DIG
Syntax :DISPlay[:WINDow]:TEXT1:CALCulate{1|2|3|4}:DIGit <numeric>
:DISPlay[:WINDow]:TEXT1:CALCulate{1|2|3|4}:DIGit?
Description Specifies the number of digits to use in the screen display of the measurement result for
each of the parameters 1 (:CALC1) through 4 (:CALC4).
Parameters
If your specified parameter is beyond the valid range, the instrument ignores the parameter
and uses either the maximum or minimum value depending on whether the parameter is
greater than the upper limit or smaller than the lower limit.
Query response {numeric}<newline><^END>
Equivalent key
sequence
[Display] - SETUP MEAS DISP - {PRMTR-1|PRMTR-2|PRMTR-3|PRMTR-4} - DIGIT
Description
ON or 1 (initial value) Turns on screen display
OFF or 0 Turns off screen display
<numeric>
Description Number of digits
Range 4 to 6
Default 4
Resolution 1