:DATA:RDC?
16. Command Reference
:DATA:RDC?
Syntax :DATA[:DATA]:RDC?
Description Returns the result of R
dc
measurement. (Query only)
Query response {numeric}<newline><^END>
If the R
dc
measurement feature is disabled, the command fails and an error is generated.
Equivalent key
sequence
No equivalent keys are available on the front panel.
:DATA:VMON?
Syntax :DATA[:DATA]:VMON?
Description Returns the result of monitoring the voltage level of the test signal. If the measurement
fails because of overload, the command returns 9.9E37. (Query only)
Query response For single-point measurement
{numeric}<newline><^END>
The command returns the monitored value (real number) of the measured point.
For list measurement
{numeric 1},...,{numeric N}<newline><^END>
The command returns the monitored value (real number) at each point defined in the active
table. Therefore, the returned data contains as many monitored values as the number of
measurement points (N) and is sent in the order of measurement point numbers.
See also :DATA:IMON? on page 263
Equivalent key
sequence
No equivalent keys are available on the front panel.