Specifications
Chapter 16 265
Command Reference
:DATA:RCAD{1-3}?
16. Command Reference
:DATA:RCAD{1-3}?
Syntax :DATA[:DATA]:RCAD{1|2|3}?
Description Returns a set of measurement data (real number) for calculating the calibration coefficients
to use during R
dc
measurement. The instrument holds three sets of calculation data, which
are obtained by measuring the OPEN, SHORT, and LOAD standards. To obtain a particular
data set, suffix the command with the number that corresponds to the desired data set. The
following table lists the command suffix numbers associated with each data set: (Query
only)
1: OPEN measurement data
2: SHORT measurement data
3: LOAD measurement data
Query response {numeric}<newline><^END>
Equivalent key
sequence
No equivalent keys are available on the front panel.
:DATA:RCCO{1-3}
Syntax :DATA[:DATA]:RCCO{1|2|3} <numeric>
:DATA[:DATA]:RCCO{1|2|3}?
Description Sets or returns a calibration coefficient for R
dc
measurement (see “Calibration coefficients
and calibration data for Rdc measurement” on page 83). To set or obtain a particular
coefficient, suffix the command with the number that corresponds to the desired
coefficient. The following table lists the command suffix numbers associated with each
coefficient:
1: Calibration coefficient A for R
dc
measurement
2: Calibration coefficient B for R
dc
measurement
3: Calibration coefficient C for R
dc
measurement
Parameters
Query response {numeric}<newline><^END>
Equivalent key
sequence
No equivalent keys are available on the front panel.
Description
<numeric> Calibration coefficient (real number) for R
dc