Specifications

Chapter 16 263
Command Reference
:DATA:FDAT{1-4}?
16. Command Reference
:DATA:FDAT{1-4}?
Syntax :DATA[:DATA]:FDATa{1|2|3|4}?
Description Returns the display data array for one of parameters 1 (:FDAT1) through 4 (:FDAT4). This
array contains the display data obtained by performing measurement parameter conversion
and data operations on the corresponding data array (see “Display Data Arrays” on
page 79).
If the measurement fails because of overload, the command returns 9.9E37. (Query only)
Query response For single-point measurement
{numeric}<newline><^END>
The command returns the value (real number) of the measured point.
For list measurement
{numeric 1},...,{numeric N}<newline><^END>
The command returns as many values (real numbers) as the measurement points defined in
the active table, in the order of measurement point numbers.
Equivalent key
sequence
No equivalent keys are available on the front panel.
:DATA:IMON?
Syntax :DATA[:DATA]:IMON?
Description Returns the result of monitoring the current level of the test signal. If the measurement
failed because of overload, the command returns 9.9E37. (Query only)
Query response For single-point measurement
{numeric}<newline><^END>
The command returns the monitored value (real number) of the measured point.
For list measurement
{numeric 1},...,{numeric N}<newline><^END>
The command returns the monitored value (real number) at each point defined in the active
table. Therefore, the returned data contains as many monitored values as the number of
measurement points (N) and is sent in the order of measurement point numbers.
See also :DATA:VMON? on page 267
Equivalent key
sequence
No equivalent keys are available on the front panel.