Specifications

262 Chapter 16
Command Reference
:DATA:CMP{1-3}
:DATA:CMP{1-3}
Syntax :DATA[:DATA]:CMP{1|2|3} <numeric 1>,...,<numeric N´2>
:DATA[:DATA]:CMP{1|2|3}?
where N represents the number of points defined in the active table.
Description Sets or returns a compensation coefficient array (see “Compensation Coefficient Arrays”
on page 81). The instrument holds three compensation coefficient arrays that correspond to
coefficients: A, B, and C. To obtain a particular array, suffix the command with the number
that corresponds to the desired array. The following table lists the command suffix numbers
associated each respective array:
1: Array that corresponds to compensation coefficient A
2: Array that corresponds to compensation coefficient B
3: Array that corresponds to compensation coefficient C
Parameters
You must break down the compensation coefficient (complex number) at each
measurement point into the real and imaginary parts so that the command receives N´2
parameters in all, where n represents an integer between 1 and N.
Query response {numeric 1},...,{numeric N´2}<newline><^END>
Equivalent key
sequence
No equivalent keys are available on the front panel.
Description
<numeric n´2-1> Real part of compensation coefficient (complex number) at nth
point
<numeric n´2> Imaginary part of compensation coefficient (complex number) at
nth point