Specifications
260 Chapter 16
Command Reference
:DATA:CCO{1-6}
:DATA:CCO{1-6}
Syntax :DATA[:DATA]:CCO{1|2|3|4|5|6} <numeric 1>,...,<numeric N´2>
:DATA[:DATA]:CCO{1|2|3|4|5|6}?
where N represents the number of points defined in the active table.
Description Sets or returns a calibration coefficient array (see “Calibration Coefficient Arrays” on
page 80). The instrument holds six calibration coefficient arrays that correspond to
coefficients: A1, B1, C1, A2, B2, and C2. To obtain a particular array, suffix the command
with the number that corresponds to the desired array. The following table lists the
command suffix numbers associated with each array:
1: Array that corresponds to calibration coefficient A1
2: Array that corresponds to calibration coefficient B1
3: Array that corresponds to calibration coefficient C1
4: Array that corresponds to calibration coefficient A2
5: Array that corresponds to calibration coefficient B2
6: Array that corresponds to calibration coefficient C2
Parameters
You must break down the calibration coefficient (complex number) at each measurement
point into the real and imaginary parts so that the command receives N´2 parameters in all,
where n represents an integer between 1 and N.
Query response {numeric 1},...,{numeric N´2}<newline><^END>
Equivalent key
sequence
No equivalent keys are available on the front panel.
Description
<numeric n´2-1> Real part of calibration coefficient (complex number) at nth point
<numeric n´2> Imaginary part of calibration coefficient (complex number) at nth
point