Specifications
Chapter 16 257
Command Reference
:CORR2:FIXT:EDEL:DIST
16. Command Reference
:CORR2:FIXT:EDEL:DIST
Syntax [:SENSe]:CORRection2:FIXTure:EDELay:DISTance <numeric>
[:SENSe]:CORRection2:FIXTure:EDELay:DISTance?
Description If you configure the instrument to use your custom test fixture connected to the test head,
this command sets the electrical length that compensates for the delay caused by the
connection of the custom test fixture.
Parameters
If your specified parameter is beyond the valid range, the instrument ignores the parameter
and uses either the maximum or minimum value depending on whether the parameter is
greater than the upper limit or smaller than the lower limit.
Query response {numeric}<newline><^END>
See also :CORR2:FIXT on page 256
Equivalent key
sequence
[Cal/Compen] - SETUP USER FIXTURE - ELEC. LENGTH
<numeric>
Description Electrical length to compensate for delay caused by fixture
connection
Range -1E6 to 1E6
Default 0
Unit m (meter)
Resolution 1E-18