Specifications

Chapter 16 255
Command Reference
:CORR2:COLL:SAVE
16. Command Reference
:CORR2:COLL:SAVE
Syntax [:SENSe]:CORRection2:COLLect:SAVE
Description Calculates the compensation coefficients and turns on the compensation mechanism. (No
query)
If you are measuring only the OPEN standard, issuing this command turns on only the
OPEN compensation feature. This is equivalent to issuing the :CORR2:COLL:OPEN
command by specifying “ON.”
If you are measuring only the SHORT standard, issuing this command turns on only the
SHORT compensation feature. This is equivalent to issuing the :CORR2:COLL:SHOR
command by specifying “ON.”
If you are measuring both the OPEN and SHORT standards, issuing this command turns on
both the OPEN and SHORT compensation features of the compensation mechanism. This
is equivalent to issuing the :CORR2:COLL:OPEN and :CORR2:COLL:SHOR
commands by specifying “ON.”
See also :CORR2:COLL on page 252
:CORR2:COLL:OPEN on page 254
:CORR2:COLL:SHOR on page 255
Equivalent key
sequence
[Cal/Compen] - COMPEN DIALOG
:CORR2:COLL:SHOR
Syntax [:SENSe]:CORRection2:COLLect:SHORt[:STATe] {ON|OFF|1|0}
[:SENSe]:CORRection2:COLLect:SHORt[:STATe]?
Description Turns on or off the SHORT compensation feature and calculates the compensation
coefficients. This command is valid only after you have obtained the SHORT standard
measurement data for calculating the compensation coefficients; otherwise, the command
fails and an error is generated.
Parameters
Query response {1|0}<newline><^END>
See also :CORR2:COLL:SAVE on page 255
:CORR2:COLL:OPEN on page 254
Equivalent key
sequence
[Cal/Compen] - COMPEN STATUS - SHORT
Description
ON or 1 Turns on the SHORT compensation feature
OFF or 0 (initial value) Turns off the SHORT compensation feature