Specifications

254 Chapter 16
Command Reference
:CORR2:COLL:OPEN
:CORR2:COLL:OPEN
Syntax [:SENSe]:CORRection2:COLLect:OPEN[:STATe] {ON|OFF|1|0}
[:SENSe]:CORRection2:COLLect:OPEN[:STATe]?
Description Turns on or off the OPEN compensation feature and calculates the compensation
coefficients. This command is valid only after you have obtained the OPEN standard
measurement data for calculating the compensation coefficients; otherwise, the command
fails and an error is generated.
Parameters
Query response {1|0}<newline><^END>
See also :CORR2:COLL:SAVE on page 255
:CORR2:COLL:SHOR on page 255
Equivalent key
sequence
[Cal/Compen] - COMPEN STATUS - OPEN
:CORR2:COLL:RF
Syntax [:SENSe]:CORRection2:COLLect[:ACQuire]:RF {STAN1|STAN2}
[:SENSe]:CORRection2:COLLect[:ACQuire]:RF?
Description Obtains the data for calculating the compensation coefficients to use during impedance
measurement performed with the specified standard. The data is obtained for only the
active table. This command is valid only when the calibration mechanism is on; otherwise,
the command fails and an error is generated. (No query)
NOTE The measurement to obtain the data is not started by execution of this command. You must
trigger the instrument after execution of this command when the trigger source is not set to
internal trigger (Int).
Parameters
See also :CORR2:COLL on page 252
:CORR2:COLL:DC on page 253
Equivalent key
sequence
No equivalent keys are available on the front panel.
Description
ON or 1 Turns on the OPEN compensation feature
OFF or 0 (initial value) Turns off the OPEN compensation feature
Description
STAN1 OPEN standard
STAN2 SHORT standard