Specifications

Chapter 16 253
Command Reference
:CORR2:COLL:DC
16. Command Reference
:CORR2:COLL:DC
Syntax [:SENSe]:CORRection2:COLLect[:ACQuire]:DC {STAN1|STAN2}
[:SENSe]:CORRection2:COLLect[:ACQuire]:DC?
Description Obtains the data for calculating the compensation coefficients to use during R
dc
measurement performed with the specified standard. The data is obtained for only the
active table. This command is valid only when the calibration mechanism is on; otherwise,
the command fails and an error is generated. (No query)
NOTE The measurement to obtain the data is not started by execution of this command. You must
trigger the instrument after execution of this command when the trigger source is not set to
internal trigger (Int).
Parameters
See also :CORR2:COLL on page 252
:CORR2:COLL:RF on page 254
Equivalent key
sequence
[Cal/Compen] - COMPEN DIALOG
Description
STAN1 OPEN standard
STAN2 SHORT standard