:CORR1:COLL:RF
:CORR1:COLL:RF
Syntax [:SENSe]:CORRection1:COLLect[:ACQuire]:RF {STAN|STAN2|STAN3|STAN4}
Description Obtains the data for calculating the calibration coefficients to use during impedance
measurement performed with the specified standard. The data is obtained for only the
active table. (No query)
NOTE The measurement to obtain the data is not started by execution of this command. You must
trigger the instrument after execution of this command when the trigger source is not set to
internal trigger (Int).
Parameters
See also :CORR1:COLL on page 244
:CORR1:COLL:DC on page 245
Equivalent key
sequence
No equivalent keys are available on the front panel.
:CORR1:COLL:SAVE
Syntax [:SENSe]:CORRection1:COLLect:SAVE
Description Calculates the calibration coefficients based on the data obtained through the
:CORR1:COLL command, or a similar command, and turns on the calibration
mechanism. You can use the :CORR1? command to check the on/off status of the
calibration mechanism.
This command is valid only after you have obtained all the data (for both impedance and
DC) necessary for all standards (OPEN, SHORT, and LOAD); otherwise, the command
fails and an error is generated. (No query)
See also :CORR1? on page 231
:CORR1:COLL on page 244
:CORR1:COLL:DC on page 245
:CORR1:COLL:RF on page 246
Equivalent key
sequence
[Cal/Compen] - CAL DIALOG
Description
STAN1 OPEN standard
STAN2 SHORT standard
STAN3 LOAD standard
STAN4 Low-loss capacitor