Specifications
244 Chapter 16
Command Reference
:CORR1:COLL
:CORR1:COLL
Syntax [:SENSe]:CORRection1:COLLect[:ACQuire] {STAN1|STAN2|STAN3|STAN4}
Description Obtains the data for calculating the calibration coefficients to use during impedance
measurement and the calibration coefficients to use during R
dc
measurement performed
with the specified standard. When a low-loss capacitor is specified as the standard, the
command obtains the data for calculating the calibration coefficient for impedance
measurement only.
The data is obtained for only the active table. (No query)
NOTE The measurement to obtain the data is not started by execution of this command. You must
trigger the instrument after execution of this command when the trigger source is not set to
internal trigger (Int).
Parameters
See also :CORR1:COLL:DC on page 245
:CORR1:COLL:RF on page 246
Equivalent key
sequence
[Cal/Compen] - CAL DIALOG
Description
STAN1 OPEN standard
STAN2 SHORT standard
STAN3 LOAD standard
STAN4 Low-loss capacitor