Specifications
242 Chapter 16
Command Reference
:CORR1:CKIT:STAN3:LIST
:CORR1:CKIT:STAN3:LIST
Syntax [:SENSe]:CORRection1:CKIT:STANdard3:LIST <numeric 1>,<numeric 2>,<numeric 3>
[:SENSe]:CORRection1:CKIT:STANdard3:LIST? <numeric 1>
Description Defines the impedance values at a specified measurement point for the LOAD standard
values of the user-defined calibration kit. To specify the parameter type to use when
defining the LOAD standard values, use the :CORR1:CKIT:STAN3:FORM command to
instruct the instrument to accept definitions based on equivalent series resistance (Rs) and
equivalent series inductance (Ls) values; equivalent series inductance (Ls) and quality
factor (Q) values: or equivalent parallel capacitance (Cp) and dissipation factor (D) values.
If you have issued the :CORR1:CKIT:LIST command by specifying “OFF” to configure
the instrument to accept fixed-value definitions for each standard of the user-defined
calibration kit, specifying a particular measurement point has no effect; regardless of the
measurement point number specified, the same value is applied to all measurement points.
Similarly, when you issue this command in its query form, the command returns the same
value regardless of the value you assign to a particular measurement number. Nevertheless,
you must specify a valid measurement point number (for example, “1”) so that the
command can successfully execute.
Parameters • Definitions based on Rs and Ls values
• Definitions based on Ls and Q values
<numeric 1> <numeric 2> <numeric 3>
Description Measurement point number Rs Ls
Range 1 to number of points (up to
32)
-1E6 to 1E6 -1E6 to 1E6
Default 1 50 0
Unit None W (ohm) H (henry)
Resolution 1 1E-18 1E-18
<numeric 1> <numeric 2> <numeric 3>
Description Measurement point number Ls Q
Range 1 to number of points (up to
32)
-1E6 to 1E6 -1E6 to 1E6
Default 1 0 0
Unit None H (henry) None
Resolution 1 1E-18 1E-18