Specifications
Chapter 16 241
Command Reference
:CORR1:CKIT:STAN3:FORM
16. Command Reference
:CORR1:CKIT:STAN3:FORM
Syntax [:SENSe]:CORRection1:CKIT:STANdard3:FORMat {RL|LQF|CDF}
[:SENSe]:CORRection1:CKIT:STANdard3:FORMat?
Description Defines the parameter type to use when defining the values specific to the LOAD standard
of the user-defined calibration kit. When the parameter type is changed, the LOAD
standard values are reset to the initial values. Therefore, define the LOAD standard values
after the parameter type is defined by this command.
Parameters
Query response {RL|LQF|CDF}<newline><^END>
See also :CORR1:CKIT:STAN3:LIST on page 242
Equivalent key
sequence
[Setup View] (Calibration kit setup display)
Description
RL (initial value) Instructs instrument to accept definitions based on equivalent
series resistance (Rs) and equivalent series inductance (Ls)
values
LQF Instructs instrument to accept definitions based on equivalent
series inductance (Ls) and quality factor (Q) values
CDF Instructs instrument to accept definitions based on equivalent
parallel capacitance (Cp) and dissipation factor (D) values