Specifications

Chapter 16 235
Command Reference
:CORR1:CKIT:STAN1:LIST
16. Command Reference
:CORR1:CKIT:STAN1:LIST
Syntax [:SENSe]:CORRection1:CKIT:STANdard1:LIST <numeric 1>,<numeric 2>,<numeric 3>
[:SENSe]:CORRection1:CKIT:STANdard1:LIST? <numeric 1>
Description Defines the equivalent parallel conductance (G) and equivalent parallel capacitance (Cp)
values at a specified measurement point for the OPEN standard of the user-defined
calibration kit.
If you have issued the :CORR1:CKIT:LIST command by specifying “OFF” to configure
the instrument to accept fixed-value definitions for each standard of the user-defined
calibration kit, specifying a particular measurement point has no effect; regardless of the
measurement point number specified, the same value is applied to all measurement points.
Similarly, when you issue this command in its query form, the command returns the same
value regardless of the specified measurement point number. Nevertheless, you must
specify a valid measurement point number (for example, “1”) so that the command can
successfully execute.
Parameters
If your specified parameter is beyond the valid range, the instrument ignores the parameter
and uses either the maximum or minimum value depending on whether the parameter is
greater than the upper limit or smaller than the lower limit.
Query response {numeric 2},{numeric 3}<newline><^END>
See also :CORR1:CKIT:LIST on page 232
:CORR1:CKIT:STAN2:LIST on page 238
:CORR1:CKIT:STAN3:LIST on page 242
Equivalent key
sequence
[Setup View] (Calibration kit setup display)
<numeric 1> <numeric 2> <numeric 3>
Description Measurement point
number
GCp
Range 1 to number of
measurement points
-1E6 to 1E6 -1E6 to 1E6
Default 1 0 8.2E-14
Unit None S (siemens) F (farad)
Resolution 1 1E-18 1E-18