Specifications
Chapter 16 233
Command Reference
:CORR1:CKIT:STAN1:DC
16. Command Reference
:CORR1:CKIT:STAN1:DC
Syntax [:SENSe]:CORRection1:CKIT:STANdard1:DC <numeric>
[:SENSe]:CORRection1:CKIT:STANdard1:DC?
Description One of the commands for setting up the OPEN standard of the user-defined calibration kit.
Use this command to define the conductance value to use during R
dc
measurement.
Parameters
If your specified parameter is beyond the valid range, the instrument ignores the parameter
and uses either the maximum or minimum value depending on whether the parameter is
greater than the upper limit or smaller than the lower limit.
Query response {numeric}<newline><^END>
See also :CORR1:CKIT:STAN2:DC on page 236
:CORR1:CKIT:STAN3:DC on page 239
Equivalent key
sequence
[Setup View] (Calibration kit setup display)
<numeric>
Description Conductance value
Range -1E6 to 1E6
Default 0
Unit S (siemens)
Resolution 1E-18