Specifications
Chapter 16 231
Command Reference
:CORR1?
16. Command Reference
:CORR1?
Syntax [:SENSe]:CORRection1[:STATe]?
Description Returns the on/off status of the calibration mechanism.
You cannot directly turn on or off the calibration mechanism. Instead, the calibration
mechanism is automatically turned on when you issue the :CORR1:COLL:SAVE
command after measuring the data required for calculating calibration coefficients. (Query
only)
Query response {1|0}<newline><^END>
See also :CORR1:COLL:SAVE on page 246
Equivalent key
sequence
No equivalent keys are available on the front panel.
:CORR1:CKIT
Syntax [:SENSe]:CORRection1:CKIT {DEFault|USER}
[:SENSe]:CORRection1:CKIT?
Description Configures the instrument to use the specified calibration kit to obtain the data necessary
for calculating calibration coefficients. You can specify the 4287A Option 007 (16195B 7-
mm calibration kit) or your custom calibration kit.
Parameters
Query response {DEF|USER}<newline><^END>
Equivalent key
sequence
[Setup View] (Calibration kit setup display)
Description
1 Calibration mechanism is on
0 Calibration mechanism is off
Description
DEFault (initial value) Configures instrument to use default 7 mm calibration kit
USER Configures instrument to use user-defined calibration kit