Specifications

Chapter 16 227
Command Reference
:CALC:EXAM:SIZE
16. Command Reference
:CALC:EXAM:SIZE
Syntax :CALCulate[:MATH]:EXAMine[:MEMory]:SIZE <numeric>
:CALCulate[:MATH]:EXAMine[:MEMory]:SIZE?
Description Sets the maximum number of measurement data logged in the volatile memory for
statistical analysis. The maximum number must be set as Param ´ Nop ´ Times where:
Param Number of displayed measurement items covering measurement
parameters 1 through 4, test signal current/voltage level and R
dc
Nop Number of measurement points in one measurement (that is, 1 during
single-point measurement or the number of measurement points in the
active table during list measurement)
Times Number of measurements needed for statistical analysis
Parameters
If your specified parameter is beyond the valid range, the instrument ignores the parameter
and uses either the maximum or minimum value depending on whether the parameter is
greater than the upper limit or smaller than the lower limit.
If your specified parameter is not a multiple of (Param ´ Nop), the instrument ignores the
parameter and uses the next highest acceptable value above the specified parameter.
Query response {numeric}<newline><^END>
Equivalent key
sequence
[Trigger Mode] - DATA LOGGING - MAX LOG SIZE
<numeric>
Description Maximum number of measurement data
Range 1 to 2,400,000
Default 3000
Resolution Obtained data number at one measurement (Param ´ Nop)