:CALC:EXAM:GET?
16. Command Reference
:CALC:EXAM:GET?
Syntax :CALCulate[:MATH]:EXAMine:GET? <numeric 1>,<numeric 2>
Description Performs statistical processing at the specified measurement point for the specified
measurement item and then returns the results. (Query only)
The number assigned to a measurement item is used to specify it. The number is assigned
to only displayed items according to the following order.
• Measurement parameter 1
• Measurement parameter 2
• Measurement parameter 3
• Measurement parameter 4
• Test signal current level monitor
• Test signal voltage level monitor
•R
dc
For example, when the measurement parameter 1, the test signal current level, and the R
dc
are displayed and the rest of the items are not displayed, the correspondence between the
numbers and the measurement items is as follows.
• 1: Measurement parameter 1
• 2: Test signal current level monitor
•3: R
dc
When the measurement data is logging to the volatile memory (RAM) at each
measurement (that is, when the logged data size has not yet reached maximum), the data
logging for statistical analysis is interrupted by execution of this command.
This command is valid only when the RAM contains measurement data for statistical
analysis (that is, when the :CALC:EXAM:STAR command is not executed); otherwise,
the command does not return query response and an error is generated.
If the obtained data number during a measurement is changed by a change in
list/single-point measurement or by a change in the number of displayed measurement
items after data logging is started, this command returns an invalid result and an error may
be generated.
Parameters
If your specified parameter is beyond the valid range, the instrument ignores the parameter
and uses either the maximum or minimum value depending on whether the parameter is
greater than the upper limit or smaller than the lower limit.
<numeric 1> <numeric 2>
Description Measurement item number
Measurement point number
*1
Range 1 to number of displayed
measurement items (up to 7)
1 to number of points (up to 32)
Resolution 1 1