Specifications
224 Chapter 16
Command Reference
:CALC:COMP:RDC:LIM
:CALC:COMP:RDC:LIM
Syntax :CALCulate:COMParator:RDC:LIMit <numeric 1>,<numeric 2>
:CALCulate:COMParator:RDC:LIMit?
Description Sets the limit range for the R
dc
limit evaluation. The limit range is inclusive of both upper
and lower limit values.
Parameters
If your specified parameter is beyond the valid range, the instrument ignores the parameter
and uses either the maximum or minimum value depending on whether the parameter is
greater than the upper limit or smaller than the lower limit.
Query response {numeric 1}, {numeric2}<newline><^END>
The returned data always lists the lower and upper limits in that order.
See also :SOUR:LIST:RDC on page 296
Equivalent key
sequence
[Setup View] (Bin sort setup display)
<numeric 1> <numeric 2>
Description
Lower limit value
*1
Upper limit value
*1
Range -1E20 to 1E20 -1E20 to 1E20
Default 0 10000
Unit W (ohm) W (ohm)
Resolution 1E-20 1E-20
*1. If <numeric 1> is greater than <numeric 2>, the instrument uses <numeric 1> as the
upper limit and <numeric 2> as the lower limit.