Specifications
Chapter 16 223
Command Reference
:CALC:COMP:OGB
16. Command Reference
:CALC:COMP:OGB
Syntax :CALCulate:COMParator:OGBins <numeric>
:CALCulate:COMParator:OGBins?
Description Specifies the boundary between good and bad bins (the OUT_OF_GOOD_BINS line) by
specifying the number of the good bin that is adjacent to the boundary. All bin numbers
that are equal to or smaller than the specified bin number indicate good bins while those
larger than the specified number represent bad bins.
NOTE When a DUT is sorted into a bad bin, the handler interface output signal
/OUT_OF_GOOD_BINS becomes active (Low). Also, the /OUT_OF_GOOD_BINS
signal becomes active when a DUT is not sorted into any of the bins 1 through 13.
Parameters
If your specified parameter is beyond the valid range, the instrument ignores the parameter
and uses either the maximum or minimum value depending on whether the parameter is
greater than the upper limit or smaller than the lower limit.
Query response {numeric}<newline><^END>
Equivalent key
sequence
[Alt] - [2] - [7] - (when the bin sort setup display is selected by [Setup View])
<numeric>
Description Number of good bin to be adjacent to the boundary
Range 1 to 13
Default 10
Resolution 1